Fluke Biomedical ULT800 User Manual
Page 15

Ultrasound Transducer Leakage Current Tester
Controls and Indicators
5
Table 2. Controls and Indicators
Item Description
READY
Glows amber when the self-test procedure
completes (approx. 8 sec.). The ULT800 is
then ready for testing.
PASS
Glows green when either the Conductivity
Test or Leakage Test passes.
FAIL
Glows red when either the Conductivity Test
or the Leakage Test fails. Pulses red when
the Leakage Test results in less than 20
μ
A,
indicating a possible open circuit condition
with invalid test results.
LOW BAT
Flashes red to indicate that the battery
requires replacement.
LEAKAGE/CONDUCTIVITY
SWITCH
Selects the test to perform.
ON/TEST BUTTON
Turns the ULT800 on and initiates the
selected test.
BATTERY COMPARTMENT
(not shown): Holds a 9-volt alkaline battery.
The instrument automatically powers off if
you do not perform a test within
12 seconds.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)