Ult800, Users guide, Figure 3. disinfection/storage tube setup – Fluke Biomedical ULT800 User Manual

Page 20

background image

ULT800

Users Guide

10

TEE TRANSDUCER
LEAKAGE CURRENT TESTER

ULT800

INSTRUCTIONS:

1. Connect the transducer and the adapters to the ULT 800.

2. Press the ON/TEST button and wait for the READY light.

3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.

4. Select LEAKAGE and press the ON/TEST button again-

observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.

ON

READY

PASS

FAIL

LOW BAT

TEST

CONDUCTIVITY

LEAKAGE

Biomedical

Dual Conductivity
Electrode with
600/213 Extenders

Disinfection
and Storage
Tube

Transducer-Specific Adapter

Fluke
ULT800

X

ecy03f.eps

Figure 3. Disinfection/Storage Tube Setup