Ult800, Users guide, Figure 3. disinfection/storage tube setup – Fluke Biomedical ULT800 User Manual
Page 20

ULT800
Users Guide
10
TEE TRANSDUCER
LEAKAGE CURRENT TESTER
ULT800
INSTRUCTIONS:
1. Connect the transducer and the adapters to the ULT 800.
2. Press the ON/TEST button and wait for the READY light.
3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.
4. Select LEAKAGE and press the ON/TEST button again-
observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.
ON
READY
PASS
FAIL
LOW BAT
TEST
CONDUCTIVITY
LEAKAGE
Biomedical
Dual Conductivity
Electrode with
600/213 Extenders
Disinfection
and Storage
Tube
Transducer-Specific Adapter
Fluke
ULT800
X
ecy03f.eps
Figure 3. Disinfection/Storage Tube Setup
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)