Fluke Biomedical 35080M User Manual
Page 58
Appendix
The Model 35050AT Dosimeter Measurements
B
B-3
(
)
offset
Zero
offset
Zero
kVp
CAL
Actual
offset
Zero
kVp
CAL
Expected
offset
Zero
kVp
Measured
kVp
Corrected
+
−
−
−
=
⎟
⎠
⎞
⎜
⎝
⎛
Example: For the Wide Range Filter Pack (37617) (50-150kVp), the Expected CAL kVp is given as 80,
and the Zero offset value is 40 kVp. However, a Model 35080M kVp Divider in CAL mode with this filter
pack inserted is connected to a Model 35050AT Dosimeter, and a kVp of 81.2 is measured. When placed
in an X-ray beam in RUN mode, the Model 35050AT Dosimeter reads 91.5 kVp. The Corrected kVp for
this exposure is calculated from the previous equation.
(
)
kVp
90
=
kVp
40
kVp
40
-
kVp
81.2
kVp
40
-
kVp
80
kVp
40
-
kVp
91.5
=
kVp
Corrected
+
⎟
⎠
⎞
⎜
⎝
⎛
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)