4 operating instructions – Fluke Biomedical 35080M User Manual
Page 24
Cadmium K-edge Mammo Filter Pack Option (37355)
Theory of Operation
4
4-3
Figure 4-2.
The kVp Correction Curves for the Cadmium K-edge Mammo Filter Pack (37355)
4.4 Operating Instructions
1. Measurement Setup - Place the Model 35080M kVp Divider in the X-ray beam on the compression
paddle or other suitable support with the axis of the kVp Divider perpendicular to the axis of the X-
ray tube. Raise the kVp Divider (with the Cadmium K-edge Mammo Filter Pack, 37355) as close as
possible to the X-ray tube. The X-ray beam should cover the entire filter pack. (See Figure 4-3.)
Figure 4-3. K-edge Measurement Setup
Minimum
Model 35080M
kVp Divider
kV
p
Indicated
Cadmium K-edge Mammo Filter
Pack (37355) kVp Correction
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)