beautypg.com

5 geometry of setup – Fluke Biomedical 35080M User Manual

Page 14

background image

Operation

Oscilloscope Zero and Offsets

2


2-3

1A)

1B)

120 kV

160
kV

110 kV

150
kV

100 kV

140
kV

90 kV

130
kV

80 kV

CAL at 80

120
kV

70 kV

110
kV

60 kV

100
kV

50 kV

90 kV

40 kV

ZERO at 40

80 kV

CAL at 80

Sensitivity 10 kV/div

Scope setting 0.1 V/div

Sensitivity 10 kV/div

Scope setting 0.1 V/div

Figure 2-1A and 1B.

Suggested Oscilloscope Settings for Use with the Wide Range Filter Pack

(37617)

As an alternative to reading directly in kV, the output of the Model 35080M kVp Divider may be read
directly in volts. The formula below is used to calculate the kV:

kV = Model 35080M voltage x 100 + ZERO kV

Example: A peak voltage of 0.300 volts is read. The ZERO kV of the Wide Range Filter Pack (37617) is
40 kV. Therefore:

kV = 0.300 x 100 + 40

= 30 + 40

= 70 kV

2.5 Geometry of Setup

The filter pack is aligned on the beam center, with the long axis of the Model 35080M kVp Divider
perpendicular to the X-ray tube axis, and the beam normal to the tabletop (see Figure 2-2). The X-ray
beam must be large enough to cover the active area of the filter pack. Distance is not critical, but a 22-
inch distance from focal spot to kVp divider is suggested.

The kVp divider must be inverted for under-table X-ray head evaluation (refer to Figure 2-3). Again, the
filter pack should be on beam center, perpendicular to the axis of the X-ray tube and the beam normal to
the table. The beam area must be large enough to cover the active area of the filter pack.