Linear Mammo Filter Pack Option (37351)
Operating Instructions
5
5-5
5. Measure other kVp values as desired. Record the LINEAR VOLTAGE, multiply by 100 and this will
be your uncorrected LINEAR KVP. Lastly, add or subtract the CORRECTION VALUE and record
the results. This now has the true kVp value for all kVp's measured.
See Appendix A for step-by-step instructions
on using the Linear Mammo Filter Pack (37351)
in conjunction with the Cadmium K-edge
Mammo Filter Pack (37355).
6. Subtract this kVp voltage from the true K-edge kVp value (see Section 4.4). (This will be the
CORRECTION VALUE applied to all kVp measurements made with the Linear Mammo Filter Pack,
37351.)
NOTE
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)