Operators Manual
5-2
5.4 Corrections
If the amount of molybdenum filtration were known, the correction could be made with the correction
curves in Figures 5-1 and 5-2. Figure 5-1 shows the correction required for total molybdenum filtration of
12 microns, 24 microns, and 35 microns (no correction is required for 24 microns). Figure 5-2 shows the
effect of adding 0.25 mm Al to the 24 microns of molybdenum.
Figure 5-1.
Correction Curves for 12, 24, and 35 Microns Molybdenum Filtrations
A machine with 35 microns of molybdenum and a reading of 29.0, add 1.0 kV; the corrected reading is
30.0 kVp (Refer to Figure 5-1). For 12 microns and a reading of 28.0 kVp, subtract 1.5 kV for a corrected
reading of 26.5 kV.
Indicated kVp
kV
Correct
ion
Linear Mammo Filter Pack (37351)
Correction for Molybdenum Filtration
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)