Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
3. Setup and Calibration Requirements
4
! copyright - 10/24/2001
3. Setup and Calibration Requirements
Warning!
This indicator was calibrated at the factory to weigh accurately with your system.
Additional calibration is not necessary under normal conditions.
The Short Form Setup & Calibration procedure allows you to change the SETUP and CAL
numbers of the indicator. You may want to perform this procedure if:
1. the indicator is being connected to different load cells, or
2. you want to adjust the calibration to match another scale system (chapter 4).
Before continuing, first write down the current SETUP and CAL numbers of your EZ indicator.
These numbers are displayed during the Self Test.
TO RUN THE SELF TEST
1. With the indicator already ON, press
to start the Self Test.
2. Press
to "pause" the Self Test while numbers are displayed.
3. Press
again to "resume".
SETUP #
CAL #
!
Do not attempt to calibrate the scale if the indicator is not reading stable weights. The
calibration procedure will not fix instability, inconsistencies, or flashing "RANGE"
messages.