Manual Programming Guide

continued

LOG INTERVALS

(THE eLock RECORDS IN MEMORY TEMPERATURE DATA THAT CAN BE
RETRIEVED USING LOCKVIEW. USING LOG INTERVALS, THE FREQUENCY
OF TEMPERATURE RECORDS FOR “IN RANGE” AND “OUT OF RANGE”
MEASUREMENTS CAN BE ADJUSTED)

Only Supervisors are allowed to adjust LOG INTERVALS.

1. Press “MENU.” LOGIN PLEASE will be displayed.
2. Scan valid credential or enter 4-14 digit PIN and press

“ENTER.”

3. Press “DOWN” button to highlight TEMPERATURE MENU and

press “NEXT/SELECT.”

4. Using the “UP / “DOWN” button, highlight LOG INTERVALS

and press “NEXT/SELECT.”

5. Log Intervals will appear along with

WHEN IN-RANGE and WHEN

OUT-OF-RANGE

6. Press “UP”/“DOWN” buttons to select value and +/- buttons

or numeric keypad to adjust.

7. Press “NEXT/SELECT” when done.

MUTE ALARM

(THE ALARM MUST BE TRIGGERED AND THE RECORDED TEMPERATURE
MUST BE OUTSIDE OF THE ACCEPTABLE PROGRAMMED RANGE IN ORDER
TO MAKE USE OF THIS FEATURE)

1. Press “MENU.” LOGIN PLEASE will be displayed.
2. Scan valid credential or enter 4-14 digit PIN and press

“ENTER.”

3. Press “DOWN” button to highlight TEMPERATURE MENU and

press “NEXT/SELECT.”

4. Using the “UP / “DOWN” button, highlight MUTE ALARM and

press “NEXT/SELECT.”

5. MUTED TEMPORARILY will flash. Note: THERMAL ALARM

MUTED will be displayed on the main screen.
A) After muting a triggered alarm, if the temperature returns

to the acceptable programmed range within 10 minutes,

THERMAL ALARM MUTED will no longer be displayed.

B) After muting a triggered alarm, if the temperature fails

to return to the acceptable programmed range after 10

minutes, **THERMAL WARNING** will appear and the

programmed Alarm Delay time will start. If this Alarm

Delay time expires and the temperature remains outside

of the acceptable programmed range, the alarm will be

triggered and **THERMAL ALARM** will appear on the

main screen.

Note: Choosing to MUTE an alarm that has already been MUTED will
extend the remaining mute time by another 10 minutes.

RESET ALARM

(THE ALARM MUST BE TRIGGERED AND THE RECORDED TEMPERATURE
MUST HAVE RETURNED TO THE ACCEPTABLE PROGRAMMED RANGE IN
ORDER TO MAKE USE OF THIS FEATURE)

1. Press “MENU.” LOGIN PLEASE will be displayed.
2. Scan valid credential or enter 4-14 digit PIN and press

“ENTER.”

3. Press “DOWN” button to highlight TEMPERATURE MENU and

press “NEXT/SELECT.”

4. Using the “UP / “DOWN” button, highlight RESET ALARM and

press “NEXT/SELECT.”

5. ALARM RESET will flash. Note: **THERMAL ALARM** will

no longer be displayed on the main screen.
A) **THERMAL ALARM** will continue to be displayed

indefinitely until the alarm is reset by following the above

RESET ALARM steps.

B) If the temperature does not return to the acceptable

programmed range after 10 minutes, **THERMAL

WARNING** will appear on the main screen.

If the temperature does not return to the acceptable range

within the programmed alarm delay time, the alarm will be

triggered and **THERMAL ALARM** will appear on the main

screen.

Credential Menu

ADD CREDENTIAL
Note: There are two status levels; Supervisor and User. Each

eLock can hold any combination of up to 3,000 Supervisors

and Users.

Only Supervisors are allowed to ADD Users and Supervisors.

1. Press “MENU.” LOGIN PLEASE will be displayed.
2. Scan valid credential or enter 4-14 digit PIN and Press

“ENTER.”

3. Press “DOWN” button to highlight CREDENTIAL MENU and

press “NEXT/SELECT.”

4. Highlight ADD CREDENTIAL and press “NEXT/SELECT.”
5. CREDENTIAL TO ADD will be displayed. Scan new credential

or enter new 4-14 digit PIN and Press “ENTER.”

6. Adding credential will be displayed.
7. Default status is User. To change status to Supervisor press

the + button.

8. The new User/Supervisor will be added to the next available

slot number. To change the slot number, press “UP” and then

+/- to move to next available slot; or use the numeric keypad

to enter the desired slot number and press “NEXT/SELECT.”