CompX eLock Instructions continued

Below are the FACTORY DEFAULT SETTINGS for eLocks

equipped with temperature monitoring.

NOTE:

1) Permission Levels can only be changed using LockView

software.

2) Permission Levels checked in gray scale cannot be changed.

Provided no Permission Level changes have been made

in LockView, Supervisor status is required to adjust the

following settings via manual programming at the eLock:
Reset Min/Max observations
MIN/MAX OBSERVATIONS IS THE THE HIGHEST AND LOWEST

TEMPERATURE THAT HAS BEEN RECORDED SINCE THE LAST

TIME THE eLock WAS RESET. THIS WILL NOT RESET THE UPPER

AND LOWER LIMITS FOR ALARMING.

At the eLock, press the “UP” button to view the highest

observed temperature or the “DOWN” button to view the

lowest observed temperature recording since the last reset

Set upper/lower limits
UPPER/LOWER LIMITS IS THE HIGHEST AND LOWEST

ACCEPTABLE TEMPERATURE. THE TEMPERATURE RANGE CAN

BE ADJUSTED IN 1/10TH DEGREE INCREMENTS.

The temperature range of a:
Standard Temp eLock: -40°C to 100°C (-40°F to 212°F)
Ultra Low Temp eLock: -100°C to -40°C (-148°F to -40°F)

Set alarm volume
OFF, SOFT, MED, LOUD

Enable/disable the alarm
TURN ALARM ON/OFF

Permission to “Enable/disable the alarm” requires that

permission to “Open while alarm is sounding” is also

granted.

Provided no Permission Level changes have been made

in LockView, BOTH Users* and Supervisors may adjust the

following settings via manual programming at the eLock:

Select Display Formats
HOURS: AM/PM or 24 HOUR
DAY/DATE: MON-SUN MM/DD/YY; MON-SUN MMM D ‘YY;

MON-SUN DD-MM-YY; MON-SUN DD-MMM-YY or MON-SUN

YYYY-MM-DD
TEMP: °FAHRENHEIT or °CELSIUS

Open while alarm is sounding
A VALID CREDENTIAL WILL BE ALLOWED TO OPEN AN eLock

WHILE AN ALARM IS SOUNDING.
*In order for User status credentials to be permitted to

open an eLock while an alarm is sounding, permission for

Supervisors to do so must also be granted.
Mute the alarm
THE ALARM MUST BE TRIGGERED AND THE RECORDED

TEMPERATURE MUST BE OUTSIDE OF THE ACCEPTABLE

PROGRAMMED RANGE IN ORDER SEE THIS FEATURE

Mute can be used to silence a triggered alarm for 10 minutes.

(see Manual Programming Guide for details)

NOTE:
Alarm Delay
and Alarm Time can only be adjusted via

LockView.