4 short form start-up, Hort form start – Bronkhorst FLOW-BUS interface User Manual

Page 6

background image

BRONKHORST

®

 

1.4 S

HORT FORM START

UP

 

 
All necessary settings for this module are already performed at Bronkhorst. Following the next steps carefully is the 
quickest way to get this module operational in your own FLOW‐BUS environment. 
 

START 

Connect 

Instrument 

Make operational 

Connect instrument to the bus

Make instrument operational on the bus (give it a FLOW‐BUS node‐address)

Link to operator module 

Check 

Ready 

Factory configured: 
Complete  FLOW‐BUS  systems  are  already factory  configured  and  tested  and  will  be 
directly operational on the bus. However, changes in the system or adding new modules 
could need new (free) addresses on the FLOW‐BUS. 
By LED’s and switch: 
Watch  the  LED’s  on  top  of  the  instrument  when  the  instrument  powers‐up.
If  the  green  LED  is  burning  continuously  and  the  red  LED  is  off,  everything  is  o.k.  the 
instrument  is  operational  on  the  bus.  If  the  red  LED  is  blinking  slowly,  the  FLOW‐BUS 
address  is  already  occupied.  In  that  case:  press  the  micro  push‐button  and  release  it 
after  2  seconds.  Now  the  instrument  is  operational  on  the  bus  (pressing  the  push‐
button  for  1  to  4  seconds  will  activate  the  auto  install  function,  the  instrument  will 
install itself to a free node‐address). 

Set address by the rotary switches on the side of the instrument (if present). 

By rotary switch: 

Make link to operator module (search module on the bus with the operator).
You can search for this instrument, looking for type and serial number.  

With E‐7000 module for digital instruments: 

Go to the FLOW‐BUS menu and select “OPERATE INSTR.”. Use <‐ or ‐> keys to search 
the bus for the wanted instrument. Instruments can be identified by their serial 
number. Also both LED’s on the instrument will blink turn by turn (wink) for some 
seconds.When found, press middle‐key (enter) to make connection for operation. 

With PC‐software application: 

When you use FLOWDDE or FLOWBUS.DLL (for Windows applications) all modules 
connected to FLOW‐BUS will be recognised automatically. They will be assigned to a 
channel for operation. (see FLOW‐DDE configuration menu) Other programs will 
have to get procedures for reading identification of instruments. For details, see 
FLOW‐BUS software documentation. 

Send  a  setpoint  to  the  instrument  and  check  the  measured  value.  Let  the  instrument 
warm‐up  for  30  minutes  for  best  accuracy.  Your  Mass  Flow  Meter/Controller  is  now 
ready for operation. 

 

 

Page 6   

 

 

 

 

FLOW‐BUS interface 

 

 

 