beautypg.com

2 troubleshooting hints and tips, Roubleshooting hints and tips – Bronkhorst FLOW-BUS interface User Manual

Page 25

background image

BRONKHORST

®

 

Page 25   

 

 

 

 

FLOW‐BUS interface 

 

 

 

9.17.024 

 

6.2 T

ROUBLESHOOTING HINTS AND TIPS

 

 
 

Symptom 

Possible cause 

Action 

No output signal on FLOW‐BUS interface 

Bad communication with FLOW‐BUS 

Check communication e.g. with local 
readout module or PC‐test program 

No or slow response to setpoint changes 

Slope value could be set too high (ramp 
function for setpoint signal) 

Check and/or change slope value with 
readout module or PC‐software program 

 

Control mode/setpoint source for 
instrument could point to a different 
setpoint source 

Check if correct control mode/setpoint 
source is selected for the instrument 

 

Other instrument (or PC) connected to 
FLOW‐BUS could change the setpoint 

Make sure who is in charge to change 
setpoints 

 

Polynomial factors could be wrong 

Check if right fluid has been selected and 
check polynomial factors A..D with those 
on your calibration certificate 

Inaccuracy of measurement 

Wrong fluid (and thus calibration settings) 
could be selected (up to 8 fluids possible) 

Check if correct fluid is selected for this 
instrument 

 

 

Check if powering of instrument is done 
correctly 

Analog output range is too low/high 

Wrong output range could be selected 

Check if wanted output range (analog) is 
selected 

No/bad connection to FLOW‐BUS 

Wrong cable used 
 

Check cabling 
 

 

Bad termination of RS485 bus‐line 

Check bus termination 

 

Other module(s) on the bus could cause 
trouble 

Check correct functioning from other 
module(s) on the bus (PC‐programs with 
interface modules included) 

No communication possible with PC‐
software program 

Interface settings could be wrong 

Check settings e.g. for RS232 (COM‐port 
and baud rate) 

 

Interface has bad connection to FLOW‐
BUS 

Check cabling of interface 

Red or green LED’s are blinking on top of 
the instrument 

Instrument is in a special operation mode 
or is indicating some kind of 
error/warning 

See manual 9.17.023 for more details 

No/bad reaction to manual instructions by 
means of micro push‐button switch on top 

Some actions can only be used under 
special conditions 

See manual 9.17.023 for more details 

Master‐slave controlling doesn’t function 
well 

Connection to master signal could be lost  Check if master output is given correctly 

 

Settings for this special controlling mode 
could be wrong 

Check if settings are correct (control mode, 
slave‐factor) 

Controller doesn’t function as wanted 

Settings for controller are for other 
behaviour than you want 

Under certain conditions you could run an 
auto‐tune for controller optimisation.  
(see manual 9.17.023) 

 

 

Change response factor settings for 
controller (see manual of readout/control 
module or  
PC‐software) 

Other troubles 

One or more parameter settings could be 
wrong 

Get back original factory settings by means 
of micro push‐button on top (see manual 
9.17.023) 

 

 

Contact sales representative for assistance 
if the above doesn’t help