beautypg.com

6 troubleshooting, 1 led indications, Troubleshooting – Bronkhorst FLOW-BUS interface User Manual

Page 24: Indications, 1 led

background image

BRONKHORST

®

 

Page 24   

 

 

 

 

FLOW‐BUS interface 

 

 

 

9.17.024 

6 TROUBLESHOOTING 

 

6.1 LED

 INDICATIONS

 

 

Led 

Time 

Indication 

 

 

 

Green 

off  

Continuous  Power‐off or program not running 

on  

Continuous  Normal running/operation mode 

Short 
flash 

0.1 sec on 

2.0 sec off 

Initialization mode 
Secured parameters can be changed 
Remote install to FLOW‐BUS 

normal 
flash 

0.2 sec on 

0.2 sec off 

Special function mode 
Instrument is busy performing any special function. 
E.g. auto‐zero or self‐test 

 

 

 

Red 

off 

Continuous  No error 

Short 
flash 

0.1 sec on 

2.0 sec off 

Node occupied: 
Re‐install instrument 

normal 
flash 

0.2 sec on 

0.2 sec off 

Warning message. 
An error occurred of minor importance. 
It would be wise to investigate the cause of this. 
You are still able to work with your instrument. 

on 

Continuous  Critical error message. A serious error occurred in the instrument. 

Instrument needs service before further using. 

Wink Mode 

 

 

 

Green 

 

 

 

Red 

 

 

 

Green 

 

 

 

Red turn by turn 

slow 
wink 

0.2 sec on 

0.2 sec off 

Wink mode 
By a command send via FLOW‐BUS the instrument can “wink” with Led’s to indicate its 
position in a (large) system 

normal 
wink 

1.0 sec on 

1.0 sec off 

Alarm indication: minimum alarm, limit/maximum alarm; power‐up alarm or limit exceeded or 
batch reached. 

fast 
wink 

0.1 sec on 

0.1 sec off 

Switch‐released, selected action started