beautypg.com

Monroe Electronics Charge Plate - Monitor - model 288B User Manual

Page 4

background image

4

Section 1

Specifications

All specifications are referred to plate voltage unless otherwise specified.



Display

240 x 64 character/graphic

Voltage -

3½ digit display (Decay and Peak reading)

Accuracy

±0.1% of reading ± 3 lsd

Resolution

1 volt for readings > 99 volts
0.1 volt for reading < 100 volts


Time -

4 digit display

Accuracy

0.1% of reading ± 1 lsd

Resolution

0.1 second for readings < 1000 seconds
1 second for reading > 999 seconds


Electrometer
Dynamic range

±1200 volts

Follower error

< 10 mV

Speed of

Response

< 10 msec for 1 kV to 0 volts (90% - 10%)

Bandwidth

-3db @ 1Khz 20V

p-p

-3db @ 10Hz 2000V

p-p

Noise

<12 mV

rms


Monitor output

Divide by 200

Accuracy

0.1% of reading ± 12 mV

Output
Impedance

1 kΩ


Start Voltages

1000 volts ±0.3% Standard

Range

±10 to ±1000 volts.

Resolution

settable to 1 volt

Accuracy

0.3% of setting ± 2.5 volts


Stop Voltages

100 volts ±3% Standard

Range

0 - ±995 volts

Resolution

settable to 1 volt

Accuracy

0.3% of setting ± 2.5 volts


Charge voltage

Range

10 to 100 volts above the start voltage

Resolution

settable to 1 volt increments

Accuracy

0.3% of setting ± 2.5 volts

This manual is related to the following products: