Casella CEL Sampling heads for personal dust sampling User Manual
Page 20

7.
SAMPLING
7.1
General Sampling Considerations
Before sampling, ensure that the sampling pump is fully charged to enable an
8-hour run time, especially for filters giving a high initial pressure drop.
The Apex and Apex Pro pumps are designed to work at a maximum
efficiency for flow rates of between 0.5 and 5 l/min.
The sampling head should be worn as near to the nose and mouth as
possible, as shown in Figure 19. It can be clipped to the OUTSIDE of a breast
pocket or to the lapel of a jacket. A harness is available to hold the equipment
in the correct position.
Note that the sampler should never be run without a filter on the inlet.
7.2
Static Sampling
Static or fixed-point sampling is possible using a personal pump and sampling
head. It will only give a rough estimate of total inhalable or respirable dust.
Note that a suitable support stand is required when performing static
sampling with a personal pump,
Sampling
Figure 19: Wearing a sampling head and Apex pump
02071
Page 20 of 24
SAMPLING HEADS AND ACCESSORIES
Users Guide