beautypg.com

Maintenance – ABUS Technologies A1500 Universal Process Indicator User Manual

Page 16

background image

ABUS TECHNOLOGIES INC.

16

A1500

9. MAINTENANCE



9.1 Troubleshooting

Connection errors or improper configuration will result in malfunctioning of the indicator.

Carefully revise all cable connections and programming parameters before operating the unit. Some

error messages will help the user identify possible problems.

MESSAGE

POSSIBLE PROBLEM

Measured value is above the programmed sensor or input signal limit.

Measured value is below the programmed sensor or input signal limit.

Open input. No sensor is connected or the sensor is broken.

Pt100 cable resistance is too high or the sensor is badly connected.

Different messages other than the ones above should be reported to the manufacturer. Please

inform the serial number if this should occur. The serial number can be viewed at the display by

pressing the

key for about 3 seconds. The software version of the instrument can be viewed at the

time the unit is powered. The instrument might display false error messages especially concerning the

type of input selected.


9.2 Special Recommendations

Should the indicator be repaired, some special handling care should be taken. The device must

be withdrawn from the case and immediately placed in an anti-static wrap; protected from heat and

humidity.


9.3 Input Calibration


Should calibration of some scale be necessary, proceed as it follows:

1. Program the indicator with the type of input requiring calibration;
2. Program the high and low limits of the measure (in.lol and in.kil) for the extreme of the type of

input programmed;

3. Assign the input a corresponding signal and a know indication/measure and slightly over the low

limit of the indication/measure;

4. Access the “

inLo

“parameter. Use the MIN and MAX keys to select the expected values;

5. Assign the input a corresponding signal and a known indication and slightly below the higher limit of

the indication/measure;

6. Access the “

inki

“parameter. Use the MIN and MAX keys to select the expected values;

7. Repeat steps c to f until no new adjustment is necessary.

Note: When verifications are preceded, note if the Pt100 excitation/activation current the calibrator

requires is compliant to the Pt100 excitation current used in this instrument: 750µA.