beautypg.com

Periodic pressure tests, Technical support, 4590 micro reactors – Parr Instrument Series 4590 User Manual

Page 17

background image

4590 Micro Reactors

w w w . p a r r i n s t . c o m

17

Periodic Pressure Tests

Each cylinder used in a Parr stirred reactor is tested
under hydrostatic pressure to the higher of 1.43
times the rated working pressure at room tem-
perature or 1.30 times the rated working pressure
corrected for temperature before it is released from
the factory. Micrometer caliper measurements are
taken during this test to check the deflection of the
walls under pressure. Excessive deflection or failure
of the metal to resume its original dimensions after
pressure is released indicates that a cylinder is po-
tentially unsafe and it will be rejected. Similar tests
should be made at regular intervals during the life
of each cylinder, and particularly whenever the user
suspects that the equipment has been over-stressed
or damaged.

Some laboratories maintain hydraulic test facilities
and make it a rule that all pressure vessels must
be tested at regular intervals. Records are kept of
deflections at specific test pressures so that any
increase in deflection becomes a warning that the
metal has lost strength. Any cylinder that fails to
return to its original dimensions after application of
the prescribed hydrostatic test should be discarded
as unsafe for further use.

Users who do not have pressure test facilities can
return any Parr pressure vessel to the factory for hy-
drostatic testing and overhaul. This should be done
whenever the metal shows excessive damage from
corrosion or whenever an over-pressure or other
unusual occurrence raises any safety questions.

To return a vessel for repair, contact Parr Instrument

Company for a return authorization number (RMA).

Apparatus returned for testing and overhaul should
be shipped prepaid to the following address:

Ship repair to:
Parr Instrument Company
Attn: Service Department
RMA # XXXXXX
211 53rd Street
Moline, Illinois 61265

An order or letter of instructions should be mailed to
the same address, as no repair work will be started
without specific instructions and a Health & Safety
Assurance Certification form (FX004) signed by a
responsible user.

Technical Support

Parr Instrument Company strives to provide our
customers with world class support. When contact-
ing Parr for technical support, please provide the
complete serial number etched on the side of the
cylinder. Serial numbers are a long string of letters
and numbers. See highlighted text in figure below
for an example of a complete serial number.

PARR INSTRUMENT COMPANY

MOLINE, IL. USA

2430HC3 222315 T316 112409 4600-1003-14106

MAWP 1900 PSI AT 350C 2010

Customer Service

Questions concerning the installation or operation

of this instrument can be answered by the Parr

Customer Service Department:

1-309-762-7716 • 1-800-872-7720

Fax: 1-309-762-9453

E-mail: [email protected]

www.parrinst.com

Each Parr operating instruction manual binder
includes a customized Parts List that contains a list
of the parts in the major components of your Parr
Reactor. Do not discard this Parts List. You should
use this parts list for reordering consumable and
replacement parts to ensure that you order the cor-
rect items for your reactor system.