Chapter 5, Reports, Chapter 5 46 – Parr Instrument 6400 User Manual

Page 48: Hapter

background image

5

Reports

P a r r I n s t r u m e n t C o m p a n y

46

c

haPter

5

Reports

Reports

The 6400 Calorimeter can transmit its stored test
data in either of two ways. The AUTO REPORT
DESTINATION
key on the Reporting Controls Menu
toggles the report destination between the display
and an optional printer connected to the USB port
of the calorimeter. This menu also selects the type
of reports that are generated automatically by the
calorimeter.

There are two kinds of reports: Preliminary and Final.

Preliminary Reports are generated at the conclu-

sion of a test. They are intended to confirm to
the operator that the results of the test fell within
the expected range.

Final reports are generated once all of the ther-

mochemical corrections have been entered into
the file. If fixed corrections are used for all of the
thermochemical corrections, a preliminary report
will not be generated. The report will automati-
cally become finalized.

Thermochemical corrections are entered by using
the following steps to select and edit preliminary
reports. Test results are stored as files using the
sample ID number as the file name. A listing of the
stored results is accessed by pressing the REPORT
command key. The REPORT command key brings
up a sub-menu on which the operator specifies.

Select From List. This key displays the stored

results specified with the following two keys.

Run Data Type. This key enables the operator to

display only determination runs, only standard-
ization runs, only solution runs (not applicable to
the 6400) or all runs.

Run Data Status. This key enables the operator

to display only preliminary reports, only final
reports, only pre weighed samples reports, all
stored reports or preliminary and final reports.

Prompt For Final Values. When turned on, the

controller will prompt the operator to enter any
missing corrections for fuse, sulfur and acid in
any selected preliminary reports. When turned off
preliminary reports will be displayed as entered.

The displayed files can be sorted by sample ID
number, by type, by status or by date of test by
simply touching the appropriate column. Individual
files can be chosen by highlighting them using the
up and down arrow keys to move the cursor. Press
the SELECT key to actually enter the selection. Once
selected the highlight will turn from dark blue to
light blue. A series of tests can be selected by scroll-
ing through the list and selecting individual files.
The double up and down keys will jump the cursor
to the top or bottom of the current display. If a range
of tests is to be selected, select the first test in the
series, scroll the selection bar to the last test in the
series and press EXTEND SEL to select the series.

The DESEL ALL key is used to cancel the current se-
lection of files. To bring the selected report or series
of reports to the display, press the DISPLAY key. To
send the reports to the printer press the PRINT key.

The EDIT key brings up a sub-menu which enables
the operator to edit any of the data in the report or
add thermochemical corrections to convert pre-
liminary reports to final reports. Final reports can
only be edited if Edit Final Reports on the Reporting
Controls Menu is turned on.

To have the Net Heat of Combustion print as part
of preliminary and final reports, go to the Thermo-
chemical Corrections Menu and then to the Net Heat/
Dry Factors Menu. Then turn ON Calculate Net Heat
of Combustion. During the reporting process, the
controller will prompt for the hydrogen (H) value.