Tables – Parr Instrument 6400 User Manual
Page 107

Tables
6400
15
w w w . p a r r i n s t . c o m
105
Determination
Fixed Fuse Correction
ON 50.0
Acid Correction
Fixed HN03 8.0
Fixed Sulfur Correction
OFF 0.0
Net Heat/Dry Factors
Net Heat & Dry Disable
Calculation Factors
Nitric Acid Factor
1.58
Acid Multiplier
0.0709
Sulfur Value is Percent
ON
Sulfur Multiplier
0.6238
Fuse Multiplier
1.0
Use Offset Correction (ISO)
OFF
Offset Value
0.0
Heat of Formation Sulfuric Acid
36.1
Heat of Formation Nitric Acid
14.1
Net Heat/Dry Factors
Fixed Hydrogen
OFF 0.0
Fixed Oxygen
ON 0.0
Fixed Nitrogen
ON 0.0
Calculate Net Heat of Combustion
OFF
Fixed Moisture as Determined
OFF 0.0
Fixed Moisture as Received
OFF 0.0
Dry Calculation
OFF
Data Entry Controls
Prompt for Bomb ID
ON
Weight Entry Mode
Touch Screen
Acid Entry Mode
Touch Screen
Net Heat Entry Modes
Auto Sample ID Controls
ON
Sample Weight Warning above
2.0
Spike Weight Entry Mode
Touch Screen
Sulfur Entry Mode
Touch Screen
Moisture Entry Modes
Auto Preweigh Controls
ON
Net Heat Data Entry Controls
Hydrogen Entry Mode
Touch Screen
Oxygen Entry Mode
Touch Screen
Nitrogen Entry Mode
Touch Screen
Moisture Data Entry Controls
MAD Entry Mode
Touch Screen
MAR Entry Mode
Touch Screen
Auto Sample ID Controls
Automatic Sample ID
ON
Automatic Sample ID Number
1
Automatic Sample ID Increment
1
Auto Preweigh Controls
Automatic Preweigh ID
ON
Automatic Preweigh ID Increment
1
Automatic Preweigh ID Number
1
Reporting Controls
Report Width
40
Automatic Reporting
ON
Auto Report Destination
Printer
Individual Printed Reports
OFF
Edit Final Reports
OFF
Recalculate Final Reports
OFF
Use New EE Values in Recalculation
OFF
Report Schedule
End of Cool/Rinse
Communication Controls
Printer Type
Parr 1758
Balance Port
Network Interface
Printer Destination
Local USB
Bar Code Port
Network Data Devices
Balance Port Communications
Balance Type
Generic
Customize Balance Settings
Balance Port Settings
Number of Data Bits
8
Parity
None
Number of Stop Bits
1
Handshaking
None
Baud Rate
9600
Data Characters from Balance
8
Data Precision
4
Transfer Timeout (seconds)
10
Balance Handler Strings
Data Logger
Data Logger
OFF
Interval in Seconds
12
Data Log Destination
Log File and Printer
Data Log Format
Text