Nor-Cal CDG 025 OP Lit User Manual
Page 22

22
CDG025-OP-LIT (2010-1)
It also permits to adjust an offset of the characteristic curve in
order to
• compensate for the offset of the measuring system or
• obtain a slightly positive zero for a 0 … 10 V AD converter.
The offset should not exceed 2% of the F.S. (+200 mV). At a
higher positive offset, the upper limit of the measurement range
is exceeded.
Recommended procedure for adjusting the offset of a
measuring system:
→ Notice 20.
Operate the gauge for at least ¼ hour (until the signal is
stable).
Push the
keep it depressed. The
5 s, the zero adjustment value, starting at the current out-
put value, keeps continually changing (ramp) until the but-
ton is released or until the setting limit (max. 25% F.S.) is
reached. The corresponding output signal is delayed by
about 1 s.
max. ø1.1 mm
Keep the button depressed