beautypg.com

Nor-Cal CDG 025 OP Lit User Manual

Page 11

background image

CDG025-OP-LIT (2010-1)

11

Materials exposed to vacuum

Flange, tube

Sensor and diaphragm

Sensor–diaphragm connection

Ceramics–metal connection

stainless steel AISI 316L

ceramics (Al

2

O

3

≥99.5%)

glass ceramics solder

AgTiCu hard solder, Vacon 70
(28% Ni, 23% Co, 49% Fe)

Internal volume

≤3.6 cm

3

Admissible pressure (absolute)

≥200 Torr (F.S.)
1 … 100 Torr (F.S.)


3 bar
2 bar

Bursting pressure (absolute)

5 bar

Admissible temperatures

Storage
Operation
Bakeout (not in operation)

–40 °C … +65 °C
+5 °C … +50 °C
≤110 °C at the flange

Relative humidity

80% at temperatures

+31 °C decreasing to 50%

at +40°C

Use

indoors only, altitude up to
2000 m NN

Degree of protection

IP 30