The output signal (measuring signal) is depending on the moun-
ting orientation. The signal difference between the vertical and
horizontal mounting orientation is:

F.S.

U / 90°

1000 Torr

≈2 mV

100 Torr

≈10 mV

10 Torr

≈50 mV

1 Torr

≈300 mV

If the gauge is operated via a controller, the zero of the

whole measuring system has to be adjusted on the
controller: first, adjust the zero of the gauge and then,
the zero of the controller.

4.2.1

<ZERO> Adjustment

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure
for zero adjustment

1000 Torr
100 Torr

20 Torr

10 Torr

2 Torr

1 Torr

<5

×10

-2

Torr

<5

×10

-3

Torr

<1

×10

-3

Torr

<5

×10

-4

Torr

<1

×10

-4

Torr

<5

×10

-5

Torr

If the final pressure in the gauge is too high for zero ad-
justment (>25% of the F.S.), the zero cannot be reached
and the LED flashes. If this is the case, activate
the factory setting and adjust the zero again (

 23).

Operate the gauge for at least 15 minutes (until the signal

is stable).