AFA 1000 / 1 -- ALARMS -- ( Velocity )
TYPE
CUSTOMER
FACTORY DEFAULTS
The AFA1000 alarm has been factory set to the listed parameter settings. To change
any parameter setting press and hold the enter button until the "Main Menu" screen is
displayed, select "Configure" to view the parameters and use the +/- and Enter buttons
to change the settings.
FOR FULL INSTALLATION / CALIBRATION / CONFIGURATION DETAILS PLEASE
REFER TO THE DATA BOOKLET PROVIDED.
Configuration Menu: -
Menu
Calibration Config
Units
Default Settings
M/SEC
ON DISPLAY
UNITS
FPM
FPM
ON
LOW AIR ALARM
FPM
60
OFF
ON LOW
AIR
CUTOFF
ON
ON
ON CUTOFF
SPEED
FPM
30
WARNING AIR ALARM
FPM
63
ON
ALARM TO WARNING AIR HYSTERESIS
FPM
3
YES
ON
HIGH AIR ALARM
NO
OFF
ON
HIGH AIR ALARM
FPM
295
ON
LOWER AIR SAMPLE FLUCTUATIONS
%
3
ON
HIGHER AIR SAMPLE FLUCTUATIONS
%
3
ON LOWER/HIGHER AIR SAMPLE DIFFERENCE
FPM
20
ON
WARN TO ALARM AIR TIME
SECONDS
5
ON
ALARM TO WARN AIR TIME
SECONDS
1
YES
ON
ON SHOW
AIR
FLOW
NO
YES
ON
SHOW TIME LINE
NO = SHOW BAR GRAPH
OFF
ENABLED
ENABLED
ON AUDIBLE
ALARM
NOT ENABLED
ON SENSOR
DIFFERENCE
%
10
Configuration Menu
NOT ACTIVE
ACTIVE ON CLOSING CONTACT
ACTIVE ON CLOSING CONTACT
ON INPUT
1
ACTIVE ON OPENING CONTACT
NONE
ALARM DISABLE
NIGHT SET-BACK
NIGHT SET-BACK
EXTERNAL ALARM
SASH HIGH
ON
INPUT 1 - FUNCTION
HIGH/LOW