Evaporation rates – Labconco CentriVa DNA Systems 7970037 User Manual
Page 35

Appendix C: CentriVap Specifications
Product Service 1-800-522-7658
31
Evaporation Rates
Solvent Tube
Size
(ml)
Number of
Samples
Sample
Size (ml)
Temp
(C)
Time to
Dry (min.)
Ballast Rate
Overall
(ml/min)
Acetonitrile
1.5 132 1 45 56 Open
2.36
1.5
132
1
60
41
Open
3.22
1.5
132
1
75
34
Open
3.88
1.5 132 1 24 30 Open
5.50
Methanol
1.5 132 1 35 82 Open
1.61
1.5 132 1 45 65 Open
2.03
1.5 132 1 60 50 Open
2.64
1.5 132 1 75 37 Open
3.57
1.5 132 1 100
28 Open
4.71
Water
1.5 132 1 45 403 Open
0.33
1.5
132
1
60
260
Open
0.51
1.5
132
1
75
188
Open
0.70
1.5
132
1
100
123
Open
1.07
Methylene Chloride
1.5 132 1 45 23 Open
5.74
1.5
132
1
60
18
Open
7.33
1.5
132
1
75
14
Open
9.42
1.5
132
1
100
12
Open
11
Ethanol
1.5 132 1 45 70 Open
1.89
1.5
132
1
60
48
Open
2.75
1.5
132
1
75
36
Open
3.67
1.5
132
1
100
25
Open
5.28