Refer to the following if your CentriVap fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance. The following failure codes may appear on the display
when problems are sensed by the internal self check routine.
CAUTION: Disconnect power before corrective action is taken.

DISPLAY ERROR CODE

CAUSE

CORRECTIVE ACTION

Heat Sensor

Sensor failure

Replace sensor assembly.

Connection failure

Repair connection.

Close Lid

Lid open

Close lid.

Latch Fail

Solenoid failure

Check component.

Switch or sensor
failure

Check connections.

Mem Fail P1

Memory failure

Push program button #1.

New memory IC chip

Push program button #1.

Bad memory IC chip

Call Labconco – Replace IC chip or
control PCB.