Kipp&Zonen UVS-E-T UV Radiometers User Manual
Page 58
58
UVIATOR software manual
20.4 Solar Zenith Angle Status Flag
The solar zenith angle status flag indicates the status of the solar zenith angle calculation.
20.5 Ozone Plugin Status Flag
The ozone plugin status flag indicates the status of determining the ozone column density using the configured ozone plugin.
20.6 Correction Factor Status Flag
The correction factor status flag indicates the status of the process of finding the correction factor for the current solar zenith
angle and ozone column density.
20.7 Radiometer Temperature Status Flag
The temperature status flag indicates the status of converting the temperature reading to actual radiometer temperature and
verifying whether the temperature is in the range as specified by the digital instrument calibration certificate.
SZA Flag
0
1
8
Description
Successful
Sun below horizon
Not calculated
Irradiance Calculation
Yes
No
No
Ozone Flag
0
1
2
8
9
Description
Successful
Interpolated
Out of range [200, 500] [DU]
Not calculated
No ozone value found
Irradiance Calculation
Yes
Yes, but ozone column density
was interpolated
Yes, but only in case the out of
range ozone column density is
supported by the UV model
No
No
Chi Flag
0
1
2
3
8
Description
Successful
E-Scie lookup failure
Chi lookup failure
Chi not found
Not calculated
Irradiance Calculation
Yes
Yes
No
No
No
Temperature Flag
0
1
2
9
Description
Successful
No temperature readings available
Temperature out of range
Temperature interpolation error (or temperature reading out of range)
Irradiance Calculation
Yes
Yes, but uncertain whether the
radiometer operated as specified
Yes, but the radiometer did not
operate as specified
Yes, but uncertain whether the
radiometer operated as specified