beautypg.com

Kipp&Zonen UVS-E-T UV Radiometers User Manual

Page 58

background image

58

UVIATOR software manual

20.4 Solar Zenith Angle Status Flag

The solar zenith angle status flag indicates the status of the solar zenith angle calculation.

20.5 Ozone Plugin Status Flag

The ozone plugin status flag indicates the status of determining the ozone column density using the configured ozone plugin.

20.6 Correction Factor Status Flag

The correction factor status flag indicates the status of the process of finding the correction factor for the current solar zenith
angle and ozone column density.

20.7 Radiometer Temperature Status Flag

The temperature status flag indicates the status of converting the temperature reading to actual radiometer temperature and
verifying whether the temperature is in the range as specified by the digital instrument calibration certificate.

SZA Flag

0
1
8

Description

Successful
Sun below horizon
Not calculated

Irradiance Calculation

Yes
No
No

Ozone Flag

0
1

2

8
9

Description

Successful
Interpolated

Out of range [200, 500] [DU]

Not calculated
No ozone value found

Irradiance Calculation

Yes
Yes, but ozone column density
was interpolated
Yes, but only in case the out of
range ozone column density is
supported by the UV model
No
No

Chi Flag

0
1
2
3
8

Description

Successful
E-Scie lookup failure
Chi lookup failure
Chi not found
Not calculated

Irradiance Calculation

Yes
Yes
No
No
No

Temperature Flag

0
1

2

9

Description

Successful
No temperature readings available

Temperature out of range

Temperature interpolation error (or temperature reading out of range)

Irradiance Calculation

Yes
Yes, but uncertain whether the
radiometer operated as specified
Yes, but the radiometer did not
operate as specified
Yes, but uncertain whether the
radiometer operated as specified