This section describes the status flags used in output files and their influence on irradiance calculation.
In general the following applies to status flags:
• a status flag value of 0 indicates
no error or successful
• higher numbers indicate
increasing severity
• a non-zero status flag does not always mean that calculations are stopped. For example, the
radiometer temperature status flag
is a warning and does not influence the calculation of irradiances.
• status flags that indicate no processing was done, such as
not calculated, occur in case prerequisites were not successful. For
example, when the sun is below the horizon, it does not make sense to calculate the correction factor and therefore determine
the ozone value. In this case the
chi flag and ozone flag will be set to not calculated.
20.1 Program Status Flag
The program status flag indicates general program status.
20.2 Radiometer Offset Status Flag
The offset status flag indicates the status of the radiometer offset voltage calculation.
20.3 Radiometer Certificate Status Flag
The radiometer certificate status flag indicates the status of the digital instrument calibration certificate the program used
for calculations.
20. Output File Status Flags
Program Flag
0
9
Description
Successful
Failed to call ozone plugin
Irradiance Calculation
Yes
No
Offset Flag
0
1
8
Description
Successful
Fallback to the radiometer offset voltage specified by the digital
instrument calibration certificate (in case of a failed automatic offset
calculation during the night).
Not calculated
Irradiance Calculation
Yes
Yes
No
Certificate Flag
0
1
9
Description
Successful
Digital radiometer calibration certificate expired
No valid digital instrument calibration certificate found
Irradiance Calculation
Yes
Yes, but recalibrate radiometer
No