Output file status flags – Kipp&Zonen UVS-E-T UV Radiometers User Manual

Page 57

background image

57

UVIATOR software manual

This section describes the status flags used in output files and their influence on irradiance calculation.

In general the following applies to status flags:
• a status flag value of 0 indicates

no error or successful

• higher numbers indicate

increasing severity

• a non-zero status flag does not always mean that calculations are stopped. For example, the

radiometer temperature status flag

is a warning and does not influence the calculation of irradiances.
• status flags that indicate no processing was done, such as

not calculated, occur in case prerequisites were not successful. For

example, when the sun is below the horizon, it does not make sense to calculate the correction factor and therefore determine
the ozone value. In this case the

chi flag and ozone flag will be set to not calculated.

20.1 Program Status Flag

The program status flag indicates general program status.

20.2 Radiometer Offset Status Flag

The offset status flag indicates the status of the radiometer offset voltage calculation.

20.3 Radiometer Certificate Status Flag

The radiometer certificate status flag indicates the status of the digital instrument calibration certificate the program used
for calculations.

20. Output File Status Flags

Program Flag

0
9

Description

Successful
Failed to call ozone plugin

Irradiance Calculation

Yes
No

Offset Flag

0
1

8

Description

Successful
Fallback to the radiometer offset voltage specified by the digital
instrument calibration certificate (in case of a failed automatic offset
calculation during the night).
Not calculated

Irradiance Calculation

Yes
Yes

No

Certificate Flag

0
1
9

Description

Successful
Digital radiometer calibration certificate expired
No valid digital instrument calibration certificate found

Irradiance Calculation

Yes
Yes, but recalibrate radiometer
No