Kipp&Zonen UVS-E-T UV Radiometers User Manual
Page 18
6.1.4 Non-stability Interpolation
Enabling
Non-stability Interpolation will instruct the program to linear interpolate the non-stability factor (ρ) between each
DCC instead of changing it abruptly between DCCs. Note that this feature is only useful for correcting UV measurements before
the last re-calibration date.
When
Non-stability Interpolation is enabled, you can also enter Radiometer Not Used periods. During such a period linear
interpolation of the non-stability is stopped and remains constant, as the radiometer does not age when it is not used. A
Radiometer Not Used period is defined is a radiometer that is not powered on and does not receive radiation. Radiometer Not
Used periods can be added and removed by pressing the Add and Remove Period buttons. Editing is initiated by double-clicking.
In order to monitor the non-stability factor, enable the
Non-stability Factor column on the Output page.
Example
18
UVIATOR software manual