beautypg.com

Calibration, Calibration principle, Calibration procedure at kipp & zonen – Kipp&Zonen CGR 4 Pyrgeometers User Manual

Page 24: Traceability to world radiometric reference, Recalibration

background image

5. Calibration

5.1. Calibration principle

An ideal radiometer gives voltage output that is proportional to the absolute irradiance level. This

relationship can be expressed as a constant ratio called ‘sensitivity’ (Sensitivity).

The sensitivity figure of a particular pyrgeometer is unique. In the case of the CGR 4 it is determined

outdoors by comparison against a reference CGR 4 pyrgeometer. The reference pyrgeometer is

calibrated outdoors regularly at the World Radiation Centre (WRC) at Davos, Switzerland.

5.2. Calibration procedure at Kipp & Zonen

CGR 4 pyrgeometers are calibrated outdoors at Kipp & Zonen under a mainly clear sky during nighttime.

The test instruments are installed next to the reference CGR 4. The pyrgeometer detector outputs (U

emf

)

and housing temperatures (T

b

) are measured each second and compressed to one minute averages.

Afterwards the downward radiation (L

d

) on the reference pyrgeometer is calculated using Formula 2.

Formula 2

For the test CGR 4’s a one minute average sensitivity is calculated using the Formula 3.

Formula 3

A final S

t

is determined using only one minute S

t

’s determined under clear sky conditions when;

Net IR-exchange

> 40 W/m².

The sum of all periods must be at least 6 hours for a successful calibration.

5.3. Traceability to World Radiometric Reference

Reference radiometers, which are calibrated annually by the World Radiation Centre in Davos, are used

for the calibration of radiometers manufactured by Kipp & Zonen. The reference radiometers are fully

characterized, i.e. linearity, temperature dependence and directional response are recorded.

Kipp & Zonen keeps two reference radiometers for each radiometer model. These reference radiometers

are sent alternate years to WRC for calibration, so production and calibration in Delft can carry on

without interruption.

5.4. Recalibration

Radiometer sensitivity changes with time and with exposure to radiation. Periodic calibration every two

years is advised. Accurate calibrations can be done outdoors under clear sky conditions by comparison

to a reference pyrgeometer of equal or higher standard.

Page 24

CGR 4 Manual

L

d

=

b

U

emf

S

+ 5.67 • 10

-8

T

4

(L

d

U

t

S

t

=

b

-

)

5.67 • 10

-8

T

4

(L

d

b

-

)

5.67 • 10

-8

T

4