beautypg.com

Maintenance – Kipp&Zonen CGR 4 Pyrgeometers User Manual

Page 15

background image

2.6. Maintenance

Once installed the pyrgeometer needs little maintenance. The dome/window must be cleaned and

inspected regularly, ideally every morning. On clear windless nights the dome/window temperature of

horizontally placed pyrgeometers will decrease, even to the dew point temperature of the air, due to

infrared radiation exchange with the cold sky. (The effective sky temperature can be 30°C lower than

the ground temperature). In this case dew, glazed frost or hoar frost can be precipitated on the top of

the outer dome and can stay there for several hours in the morning. An ice cap on the dome is a strong

diffuser and increases the pyrgeometer signal drastically, up to 100% in the first hours after sunrise.

Hoar frost disappears due to solar radiation during the morning, but should be wiped of as soon as

possible manually.

The dome of the pyrgeometer can be ventilated continuously by a heated blower to keep the

dome/window above the dew point temperature. The need for heating strongly depends upon local

climatological circumstances. Generally, heating is advised during cold seasons when frost and dew

can be expected. The Kipp & Zonen CV 2 ventilation unit is specially designed for unattended operation

under most weather conditions and has a choice of heating levels.

Note that the CGR 3 pyrgeometer cannot be used with the CV 2 ventilation unit.

A periodic check is to ensure that the pyrgeometer is level and that the silica gel desiccant is still

coloured orange. When the orange silica gel in the drying cartridge is turned completely transparent

(normally after several months), it must be replaced by fresh silica gel as supplied in the small refill

packs. The content of one pack is sufficient for one complete refill. At the same time check that the

pyrgeometer mounting is secure and that the cable is in good condition.

Some tips when changing the desiccant:

-

Make sure the surfaces of the pyrgeometer and the drying cartridge that touch the rubber o-ring

are clean (corrosion can do a lot of harm here and dirt, in combination with water, can cause this);

-

The rubber o-ring is coated with a silicon grease to improve the seal. If the rubber o-ring looks dry

apply some grease to it (Vaseline will also do);

-

Check that the drying cartridge is tightly threaded into the pyrgeometer body.

It is very difficult to make the pyrgeometers hermetically sealed; so, due to pressure differences between

the inside and the outside of the instrument, there will always be some exchange of (humid) air.

The pyrgeometer sensitivity changes with time and with exposure to radiation. Calibration every two years

is advised. Further information about Kipp & Zonen recalibration services can be found in Appendix V.

Page 15

CGR 4 Manual