For the CGR 4 the effect of each parameter on the sensitivity can be shown separately. The non-linearity

error, the sensitivity variation with irradiance, is the same for any CGR 4 and is shown in Figure 8.

Figure 8:

Non-linear sensitivity variation with irradiance of a CGR 4 pyrgeometer

The temperature dependence of the sensitivity is a function of the individual CGR 4. For a given CGR 4 the

response lies in the region between the curved lines in Figure 9. The temperature dependence of each CGR

4 pyrgeometer is characterised and supplied with the instrument. Each CGR 4 has a built-in temperature

sensor to allow corrections to be applied if required.

Figure 9:

Typical temperature dependency of a CGR 4

Page 14

CGR 4 Manual