beautypg.com

Electrical installation – Kipp&Zonen CGR 4 Pyrgeometers User Manual

Page 10

background image

However, if the upward and downward radiation components are to be measured separately it is necessary

to record the individual housing temperatures to calculate the radiation values.

Using the combination of a net pyrgeometer (two CGR 3 or CGR 4 instruments) and a CMA 6 or CMA 11

albedometer the net total radiation (energy balance) can be calculated with high accuracy from thefour

component values. Problems with dew deposition, frost, etc, can be minimised by using the Kipp &

Zonen CV 2 ventilation unit with optional heating.

This has many advantages over conventional net total radiation sensors with plastic (polyethylene)

windows. These cannot provide individual short and long-wave radiation values and cannot separate

upward and downward contributions. The soft plastic domes do not fully protect the sensor from the

thermal effects of wind and rain, are easily soiled, are difficult to clean and require regular replacement.

2.3. Electrical installation

As standard the CGR is supplied with a waterproof connector pre-wired to 10 m cable with a number of

leads and a shield covered with a black sleeve. The number of connector pins and cable leads depends

upon the type of temperature sensor that is fitted. The colour code of the wires and the connector pin

numbers are shown on the instruction sheet. Longer cables are available as options.

Preferably, secure the pyrgeometer with its levelling screws or mounting rod to a metal support with a

good connection to earth (e.g. by using a lightning conductor).

The shield of the cable is connected to the aluminium radiometer housing through the connector body.

The shield at the cable end may be connected to ground at the readout equipment. Lightning can induce

high voltages in the shield but these will be led off at the pyrgeometer and data logger.

Kipp & Zonen pyrgeometer cables are of low noise type, but bending the cable produces small voltage

spikes, a tribo-electric and capacitance effect. Therefore, the cable must be firmly secured to minimise

spurious responses during stormy weather.

The impedance of the readout equipment loads the temperature compensation circuit and the thermopile.

It can increase the temperature dependency of the pyrgeometer. The sensitivity is affected more than 0.1%

when the load resistance is less than 100 kΩ. For this reason we recommend the use of readout equipment

with an input impedance of 1 MΩ or more. The solar integrators, data loggers and chart recorders from

Kipp & Zonen meet these requirements.

Long cables may be used, but the cable resistance must be smaller than 0.1% of the impedance of the

readout equipment. It is evident that the use of attenuator circuits to modify the calibration factor is not

recommended because the temperature response will also be affected.

A high input bias current at the readout equipment can produce several micro-Volts across the impedance

of the pyrgeometer and cable. The zero offset can be verified by replacing the pyrgeometer impedance at

the readout equipment input terminals with a resistor.

The pyrgeometer can also be connected to a computer or data acquisition system. A low voltage analogue

input must be available. The resolution of the Analogue-to-Digital Converter (ADC) must allow a system

sensitivity of about 1 bit per W/m². More resolution is not necessary during outdoor measurements,

because even the best pyrgeometer (the CGR 4) exhibits offsets greater than 2 W/m² due to lack of

thermal equilibrium.

Page 10

CGR 4 Manual