beautypg.com

Sensitivity, Impedance, Response time – Kipp&Zonen CGR 4 Pyrgeometers User Manual

Page 20: Non-linearity, Temperature dependence, Tilt error

background image

4.2. Sensitivity

The radiometer thermopile sensitivity is mainly determined by the physical properties of the detector

itself. The thermoelectric power, thermal conductivity of the junctions and the overall dimensions of the

sensing element are related to its sensitivity.

4.3. Impedance

The radiometer impedance is defined as the total electrical impedance at the radiometer output connector

fitted to the housing. It arises from the electrical resistance in the thermal junctions, wires and passive

electronics within the radiometer.

4.4. Response time

Any measuring device requires a certain time to react to a change in the parameter being measured. The

radiometer requires time to respond to change in the incident radiation. The response time is normally

quoted as the time for the output to reach 95% (sometimes 63%) of the final value following a step-change

in irradiance. It is determined by the physical properties of the thermopile and the radiometer construction.

CGR series pyrgeometers have a fast response, which makes them suitable for measuring far infrared

radiation (FIR) under variable weather conditions.

4.5. Non-linearity

The non-linearity of a pyrgeometer is the percentage deviation in the sensitivity over a net irradiance

range from -250 to +250 W/m² compared to the sensitivity calibration irradiance of -100 W/m². The

non-linearity effect is due to convective and radiative heat losses at the black absorber surface which

make the conditional thermal equilibrium of the radiometer non-linear.

4.6. Temperature dependence

The sensitivity change of the pyrgeometer with ambient temperature change is related to the thermodynamics

of the radiometer construction. The temperature dependence is given as percent deviation with respect to

the calibrated sensitivity at +20°C. The CGR 4 has passive electrical compensation circuits to minimise

this effect. Each CGR 4 pyrgeometer is supplied with an individual test certificate stating the temperature

dependency in the range from -20°C to +50°C, at 10°C intervals. The CGR series pyrgeometers are fitted

as standard with an internal temperature sensor to allow sensitivity corrections to be applied if desired.

4.7. Tilt error

This is the deviation from the sensitivity at 0° tilt (exactly horizontal) over the range from 0° to 90° tilt.

The tilt response is proportional to the incident radiation. The error could be corrected for, in applications

where it is necessary to install the pyrgeometer on an inclined surface, but is usually insignificant.

Page 20

CGR 4 Manual