Accessories – Hanna Instruments HI 98188 User Manual

Page 67

background image

67

The process can be interrupted by pressing Stop or ESC key.
The replatinization process takes about 5 minutes.

Remove the probe from the replatinization solution and rinse it with deionizated
water.

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

EC CALIBRATION SOLUTIONS

HI 70033C

84

m

S/cm, 20 ml sachet, 25 pcs.

HI 70031C

1413

m

S/cm, 20 ml sachet, 25 pcs.

HI 70039C

5000

m

S/cm, 20 ml sachet, 25 pcs.

HI 70030C

12880

m

S/cm, 20 ml sachet, 25 pcs.

HI 6033

84

m

S/cm, 500 ml, bottle

HI 6031

1413

m

S/cm, 500 ml, bottle

HI 7039L

5000

m

S/cm, 500 ml, bottle

HI 7030L

12880

m

S/cm, 500 ml, bottle

HI 7034L

80000

m

S/cm, 500 ml, bottle

HI 7035L

114800

m

S/cm, 500 ml, bottle

HI 7037L

100% NaCl, 500 ml, bottle

RECOMMENDATIONS FOR USERS

Before using these products, make sure they are entirely suitable for the environment in which

they are used.

Operation of these instruments in residential areas could cause unacceptable interferences to

radio and TV equipment, requiring the operator to follow all necessary steps to correct inter-

ferences.

The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid

touching this glass bulb at all times.

During operation, ESD wrist straps should be worn to avoid possible damage to the electrode

by electrostatic discharges.

Any variation introduced by the user to the supplied equipment may degrade the instruments’

EMC performance.

To avoid electrical shock, do not use these instruments when voltages at the measurement

surface exceed 24 Vac or 60 Vcc.

To avoid damage or burns, do not perform any measurement in microwave ovens.