beautypg.com

Hanna Instruments HI 98188 User Manual

Page 60

background image

60

NSLx

Requests the number of logged samples.

The command parameter (1 char):

• E - the request is for EC range
• R - the request is for Resistivity range
• T - the request is for TDS range
• N - the requests is for NaCl range
• U - the requests is for USP range

The answer string is the number of samples (4 chars)

ULS

Requests information about the status of the USP log reports.

The answer string contains:
• the number of reports (3 chars)
For every USP report the following information are sent:

• report ID (3 chars)
• log time (12 chars)

LLS

Requests information about the number of lots

The answer string contains:
• the number of lots (3 chars)
For every lot the following information are sent:

• lot ID (3 chars)
• log time (12 chars)
• lot type (1 char): 0 - EC; 1 - Resistivity;

2 - TDS; 3 - NaCl

GLDxxx

Requests all the records for the lot with ID = xxx
The answer string contains:
Lot header data:

• logging interval (1 s) (5 chars)
• temperature source (1char)
• temperature compensation mode (1 char)
• reference temperature (2 chars)
• temperature coefficient (6 chars)
• offset factor (6 chars)
• TDS factor (6 chars) (only for TDS lots)
• salinity coefficient (7 chars) (only for salinity lots)
• profile ID (4 chars)
• profile creation time (12 chars)
• lot start time (12 chars)
Lot record data: