beautypg.com

Hanna Instruments HI 98188 User Manual

Page 12

background image

12

According to the definition, salinity of a sample in psu (practical salinity units)
is calculated using the following formula:

R

T

- coefficient;

C

T

(sample) - uncompensated conductivity at T °C;

C(35,15)= 42.914

m

S/cm - the corresponding conductivity of KCl solution

containing a mass of 32.4356 g KCl / 1 Kg solution;

r

T

- temperature compensation polynom

a

0

= 0.008

b

0

=0.0005

a

1

= -0.1692

b

1

= -0.0056

a

2

= 25.3851

b

2

= -0.0066

a

3

= 14.0941

b

3

= -0.0375

a

4

= -7.0261

b

4

= 0.0636

a

5

= 2.7081

b

5

= -0.0144

c

0

= 0.008

c

1

= 0.0005

X = 400R

T

Y = 100R

T

f (T)= (T-15)/[1+0.0162(T-15)]

Note:

• The formula can be applied for salinity values between

0 and 42 psu.

• The formula can be applied for temperatures between

-2 °C and 35 °C.

To reach this range press Mode while in Salinity range until the practical
salinity scale [PSU] is displayed.