beautypg.com

Hanna Instruments HI 2400 User Manual

Page 27

background image

27

GLDxxxfff:Requests the records of the “xxx” lot number. The records are

sent in frames of 10 records; “fff” is the frame number (01 first
frame). (Example: Lot 13 has 53 records. The records will be
sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:

xxx - Lot number (eq: for lot number 1 xxx = 001)
fff - requested frame number - first frame is labeled 01

Notes: • “Err3” is sent if the Log on demand is empty.

• “Err4” is sent if the requested set parameter is not available.
• “Err5” is sent if the command argument is wrong.
• “Err6” is sent if the requested range is not available.
• “Err7” is sent if the instrument is logging.
• “Err8” is sent if the instrument is not in measurement mode.
• Invalid commands will be ignored.