Hanna Instruments HI 2400 User Manual

Page 25

background image

25

CHR xx

Change the instrument range according with the parameter
value (xx):

• xx=06 % range

• xx=07 ppm range

The instrument will answer for these commands with:

>

where:

is 02 ASCII code character (start of text)

is 03 ASCII code character (end of text)

:

is 06 ASCII code character (recognized command)

is 21 ASCII code character (unrecognized command)

is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

The instrument will answer for these commands with:

where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.

All the answer messages are with ASCII characters.
RAS

Causes the instrument to send a complete set of readings.

The answer string contains:

• Meter mode (2 chars):

• 06 - % range

• 07 - ppm range

• Meter status (2 chars of status byte): represents a 8

bit hexadecimal encoding.

• 0x10 - temperature probe is connected

• 0x01 - new GLP data available

• 0x02 - new SETUP parameter

• Reading status (1 char): R - in range, O - over

range, U - under range.

• The reading (corresponding to the selected range) -

7 ASCII chars, including sign and decimal point.

• Temperature reading - 7 ASCII chars, with sign and

two decimal points, always in ºC.