S E T U P

S E T U P

S E T U P

S E T U P

S E T U P

Setup mode allows viewing and modifying the following parameters:
• Salinity Factor
• Altitude Factor
• Log Interval
• Current Time (hour & minute)
• Current Date (month, day & year)
• Beep Status
• Instrument ID
• Temperature Unit
To enter the Setup mode press

SETUP while the instrument is in

measurement mode. Press

SETUP again to exit SETUP mode.

Select a parameter with the

ARROW keys.

Press

CAL if you want to change a parameter value. The selected

parameter

will blink.

Press

RANGE to toggle between displayed parameters.

Press the

ARROW keys to increase or decrease the displayed value.

Press

CFM to save the modified value or CAL to escape.

Press the

ARROW keys to select the next/previous parameter.

SALINITY FACTOR
Press

CAL when the salinity factor is displayed. The salinity factor (“0” to

40” g/l) and the “CFM” tag will start blinking.

Press

ARROW keys to change the salinity factor value.

Press

CFM to save the modified value or press CAL to escape without

saving.
The salinity affects the DO concentration, decreasing its value. The next
table shows the maximum oxygen solubility at various temperatures and
salinity levels.

• The instrument ID.