Hanna Instruments HI 2400 User Manual
Page 26

26
MDR
Requests the instrument model name and firmware code
(16 ASCII chars).
GLP
Requests the calibration data record.
The answer string contains:
• GLP status (1 char): represents a 4 bit hexadecimal
encoding.
• 0x04 - DO calibration available
• DO calibration data (if available), which contains:
• the number of calibrated satandards (1 char)
• the calibration time,
yymmddhhmmss (12 chars)
• standards information
• standard value, with sign and decimal point
(7 chars).
• the standard temperature, with sign and decimal
point (7 chars)
• the salinity value (4 chars)
• the altitude value (4 chars)
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)
• SETUP information (2 chars): 8 bit hexadecimal
encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• Log type - 01 to 12 - the coresponding interval for
lot log (5 s to 180 min)
• Salinity value (4 chars)
• Altitude value (4 chars)
LLsxff: requests information about all lots and it sends the information in
frames of 10 lots each (a frame contains information about 10 lots)
Command Parameters:
x = D - request DO data
ff - requested frame number - first frame is labeled 01