ZERO CALIBRATION

• Submerse the probe into

HI 7040 zero oxygen

solution and stir gently for 2-3 minutes.

• Press

CAL. The “~” and “ ” tags will blink on the

LCD until the reading is stable.

• When the reading is stable, “

CFM” starts blinking.

Press

CFM to confirm the “0.0%” DO calibration.

• If the reading is within the limits (±15% f.s.), the

meter stores the value (and adjusts the slope point).

• Press

CAL. The instrument will return to measurement

mode and will memorize the zero calibration data.
For a two-point calibration do not press

CAL and

follow the procedure below.

SLOPE CALIBRATION

It is suggested to perform the slope calibration in air.
• Rinse the probe in clean water to remove any residual

zero oxygen solution.

Note: If you did not perform the zero calibration procedure, press CAL

and then the

ARROW keys to select the 100% DO calibration

point.

• Dry the probe tip and allow a few seconds for the LCD reading to

stabilize. The “~

~

~

~

~” and “ ” tags will blink until the reading is

stable.

• When the reading is stable, “

CFM” tag starts blinking. Press CFM to

confirm the “

100.0%” DO calibration.

• If the reading is within the limits (±15% f.s.), the

meter stores the value (and adjusts the slope point).

OR