• Submerse the DO probe in the second vessel

as near as possible to the reference thermom-
eter. Allow a few seconds for the probe to
stabilize.

• Use the

ARROW keys to set the reading on

the secondary LCD to that of the hot water.

• When the reading is stable, the “

CFM” tag starts

blinking.

• Press

CFM to confirm. The instrument returns to

measurement mode.

Note: If the reading is not close to the selected calibration point,

WRONG” tag will blink. In this case, check if the value set on the

secondary LCD and the temperature measured with reference
thermometer are close. Change the DO probe or restart calibration
if necessary.