Hanna Instruments HI 83207-01 User Manual

Page 44

background image

44

Nickel LR

• Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, wait

for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.

• Fill one cuvette up to the mark with 10 mL of the blank.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Fill a second cuvette up to the mark with 10 mL of the reacted sample.

• Insert the second cuvette into the instrument.

• Press Read to start the reading. The instrument displays the results in mg/L of nickel.

Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking
readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity
is removed.

10 mL

10 mL