Fluoride

FLUORIDE

SPECIFICATIONS

Range

0.00 to 2.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, SPADNS method. The reaction between fluoride and the liquid reagent

causes a red tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93729-0

SPADNS Reagent

4 mL

REAGENT SETS

HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 65.

MEASUREMENT PROCEDURE

• Select the

Fluoride method using the procedure described in the

Method Selection section (see page 12).

• Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.

• Fill one of the cuvettes with distilled water up to the mark,

replace the cap and invert several times to mix.

• Fill the other cuvette with sample up to the mark, replace the

cap and invert several times to mix.

• Place the cuvette with the reacted distilled water (# 1) into

the holder and close the lid.

# 1

# 1

# 2

10 mL

10 mL

2 mL