Hanna Instruments HI 83207-01 User Manual

Page 36

background image

36

Manganese LR

• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent

solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.

• Add 1 mL of the HI 93748C-0 0.1% PAN indicator

solution to each cuvette, replace the caps and shake
gently.

• Add 3 drops of Dispersing Agent (HI 93703-51) to each

cuvette, replace the caps and invert gently to mix for
about 30 seconds.

• Place the cuvette with the reacted deionized water (blank)

into the holder and close the lid.

• Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for

2 minutes and then press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.