Hanna Instruments HI 83207-01 User Manual
Page 38

38
Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range
0.0 to 40.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.3 mg/L ±5% of reading at 25 °C
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93730A-0
Reagent A
1 packet
HI 93730B-0
Reagent B
1 packet
HI 93730C-0
Reagent C
1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Molybdenum method using the procedure
described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL