Copper HR
COPPER HIGH RANGE
SPECIFICATIONS
Range
0.00 to 5.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.02 mg/L ±4% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm
Method
Adaptation of the
EPA method. The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93702-0
Bicinchoninate
1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Copper HR method using the procedure described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-”” when the meter is zeroed and ready for
measurement.
10 mL