Hanna Instruments HI 83266 User Manual
Page 20

38
39
Iron
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 3 minutes and press
Read. When the timer ends the meter will perform the reading.
• The instrument displays concentration in
mg/L of iron.
INTERFERENCES
Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10000 ppm (as CaCO
3
)
Magnesium above 100000 ppm (as CaCO
3
)
Chloride above 185000 ppm.
OZONE
SPECIFICATIONS
Range
0.00 to 2.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.02 mg/L ±3% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Colorimetric DPD Method. The reaction between ozone and the DPD reagent causes a
pink tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity/test
HI 93757-0
DPD Powder Reagent
1 packet
HI 93703-52-0 Glycine Powder (Optional Reagent) 1 packet
REAGENT SETS
HI 93757-01
Reagents for 100 tests
HI 93757-03
Reagents for 300 tests
HI 93703-52
Glycine Powder, Optional Reagent for 100 tests
For other accessories see page 46.
IMPORTANT NOTE: Chlorine is a strong interferent for ozone determination. If the sample is suspected to
contain chlorine residues (free or total chlorine), please follow the
alternative measurement procedure
described below:
• Perform the Standard Measurement Procedure and take note of the reading:
value A.
• On a fresh sample perform the Additional Measurement Procedure and take note of the reading:
value B.
• Subtract reading
B from reading A to obtain the ozone concentration in mg/L:
mg/L (O
3
) =
value A – value B.
STANDARD MEASUREMENT PROCEDURE
• Select the
Ozone method using the procedure described
in the
Method Selection section (see page 15).
• Fill the cuvette with 10 mL of unreacted sample, up to
the mark, and replace the cap.
• Place the cuvette into the holder and close the lid.
Ozone
10 mL