20

21

SPECIFICATIONS

Range

0 to 500 mg/L (as CaCO

3

)

Resolution

5 mg/L

Accuracy

±5 mg/L ±10% of reading at 25 °C

Typical EMC

±5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Colorimetric Method

. At different alkalinity levels a distinctive range of colors from yellow

to green and greenish blue will develop.

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 93755-0

Alkalinity Indicator Reagent

1 mL

REAGENT SETS

HI 93755-01 Reagents for 100 tests
HI 93755-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Alkalinity method using the procedure described in the

Method Selection section (see page 15).

• Fill the cuvette with 10 mL of unreacted sample, up to the mark,

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

ALKALINITY

Alkalinity

10 mL

• Remove the cuvette.

• Carefully add exactly 1 mL of HI 93755-0 Liquid Alkalinity

Reagent using the supplied syringe.

• Replace the cap and invert 5 times.

• Reinsert the cuvette into the instrument and close the lid.

• Press

Read to start the reading.

• The instrument displays the results in

mg/L of alkalinity (CaCO

3

).

Note: If using a meter with software version 1.14 or earlier, readings can be improved for samples

with less than 75 ppm alkalinity by adding 0.7 mL of reagent instead of 1.0 mL.