36

37

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 45 seconds and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in

mg/L of cyanuric acid.

INTERFERENCES

Turbidity preexisting in the sample causes interference during measurement.

Cyanuric Acid

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.04 mg/L ±2% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

EPA Phenantroline method 315B, for natural and treated waters.

The reaction between iron and reagents causes an orange tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 93721-0

Iron High Range Reagent

1 packet

REAGENT SETS

HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Iron method using the procedure described in the Method

Selection section (see page 15).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette and add the content of one packet of

HI 93721-0 reagent. Replace the cap and shake until dissolution is
complete.

IRON

Iron

10 mL